Descargar Imprimir esta página

MERCK Millipore Millex-FG SLFG025LS Guia Del Usuario página 4

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 15
Directions for Use
To use the device with a vacuum or air system, attach
the appropriate Luer adapters to the vacuum or air
system tubing . If the device becomes momentarily
inoperable due to aqueous fluid completely filling the
housing, clear the filter by disconnecting it from the
upstream Luer connection and aseptically applying air
pressure with a syringe .
To use the device as a syringe filter, follow the
directions below . When filtering aqueous solutions,
filter 10 mL of ethanol or methanol through the unit to
wet the hydrophobic filter . Discard the first 2–3 mL of
filtered solution to prevent alcohol contamination .
1 . Wearing gloves using aseptic
technique: fill syringe with solution to
be filtered .
2 . Remove cover from package.
3 . Attach suitable syringe to filter inlet
and remove assembly from package .
4 . Hold syringe with filter pointing up and
top off by pushing a few drops through
to wet filter .
Excess solution may be hazardous and
Y
should be disposed with care .
5 . Depress syringe plunger to filter
solution into suitable container or
compatible attachment on outlet .
PR05859w
Rev 06/23
Specifications and
Performance Characteristics
Materials
Membrane
Pore size
Housing
Dimensions
Inlet to outlet
Diameter
Filtration area
Temperature limit
Pressure limit at 21 °C
Filtration volume
Hold-up volume
Air flow rate
Sterilization method
Connections
3273IL1mac
3273IL2mac
3273IL5mac
3251IL1mac
4 of 53
Hydrophobic Fluoropore™
polytetrafluoroethylene
(PTFE), type FG
0 .2 µm
Polyvinyl chloride (PVC)
25 mm
29 mm (1 .14 in .)
4 cm
(0 .62 in
)
2
2
45 °C (113 °F)
5 .2 bar (75 psi) inlet and
differential maximum
1–100 mL
≤ 0.1 mL after air purge
≥ 1.0 L/min at 1 psi
Ethylene oxide gas
Female Luer-Lok™ inlet;
male Luer-slip outlet

Publicidad

loading