SETUP (Continued)
ASSEMBLY INSTRUCTIONS:
STEP 1: Loosen the cylinder attachment knob.
STEP 2: Lower the EVOLUTION conserver over any CGA-870 post-valve cylinder with the
alignment pins toward the holes on the cylinder neck [see Fig. B].
STEP 3: Line up the two pins and the seal washer (gasket) with the corresponding holes on
the cylinder post valve.
STEP 4: While holding the unit in place, tighten the cylinder attachment knob by turning
clockwise [see Fig. B].
NOTE: The cylinder attachment knob should be aligned with the
indentation on the post valve.
NOTE: Tighten only by hand. The use of a tool to tighten the knob may damamge the unit.
STEP 5 : A ttach a standard cannula (7 ft. [2.13 m] or less in length) to the oxygen supply outlet . See Fig. F for an illustration
demonstrating the proper positioning of the nasal cannula.
INSPECTION BEFORE EACH USE
STEP 1: Examine the outside of the cylinder and regulator assembly for debris, oils or grease and noticeable signs of
damage, dents or dings, corrosion, excessive heat or fire damage, pressure leak or any other sign of damage that
might cause a cylinder to be unacceptable or unsafe for use.
WARNING: Fire Hazard
If any of these conditions exist, DO NOT use the cylinder and contact your healthcare provider for
a replacement cylinder.
STEP 2: Check the flow selector knob to be sure it rotates and stops at each setting.
14