INTRODUCTION
This manual provides information necessary to operate the EVOLUTION
conserver.This conserver can be used with CGA 870 post-valve cylinder [see Fig. A] at home or away from home
to provide your specific oxygen requirements. It requires two (2) 1.5 volt AA alkaline batteries for operation.
PURPOSE:
TM
The EVOLUTION MOTION OM-900M electronic oxygen conserver includes a combination of a low-pressure regulator and an
oxygen conserver. It is designed for use with a cylinder as an ambulatory oxygen system and is capable of delivering a specific
amount of supplemental oxygen at the optimal point in the breathing cycle. The EVOLUTION
delivery of oxygen, maximizing the beneficial effects and eliminating unnecessary oxygen waste.
FUNCTION :
When we breathe, approximately one-third of the time is spent inhaling and
two-thirds exhaling. As a result, oxygen delivered by continuous flow is
wasted during exhalation. By eliminating oxygen flow during exhalation, a
two-thirds saving is possible. A
very first part of inhalation contributes most to meeting oxygen needs.
The OM-900 M electronic oxygen conserver takes advantage of these facts
to provide maximum efficiency in the delivery of oxygen. This device is
designed to be an integral component of a lightweight, long-lasting
supplemental ambulatory oxygen system.
The addition of the patented motion-sensing technology makes the OM-900M a
unique therapeutic device. For those patients who have increased oxygen
requirements based on their activity level, the OM-900M stores two selectable
settings; one for rest and one for activity. By sensing the user's activity, the
device can automatically switch from a lower setting in "Rest" mode to a higher
setting in "Active" mode to meet the user's increased oxygen needs.
dditionally, the oxygen available during the
TM
MOTION OM- 9 00 M electronic oxygen
TM
increases the efficiency in the
Toggle
Parallel Holes
on the
Post Valve
F IGURE A
Post-Valve Cylinder
7