Appendix A: EMC Information
Guidance and Manufacturer's Declaration- Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The
user of this device should make sure it is used in such an environment.
Emissions Test
Compliance
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Flicker emissions
IEC61000-3-3
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The
user of this device should make sure it is used in such an environment.
Immunity Test IEC60601
test level
Electrostatic
±6kV contact
Discharge (ESD)
±8kV air
IEC61000-4-2
Electrical fast
± kV for power supply
line
IEC61000-4-4
Surge
± kV line(s) to line(s)
IEC61000-4-5
±2 kV line(s) to earth
<5 % UT (>95 % dip in
Voltage dips, short
UT) for 0,5 cycle
interruptions and
40 % UT (60 % dip in
voltage variations on
UT) for 5 cycles
power supply input
70 % UT (30 % dip in
lines
UT) for 25 cycles
IEC61000-4-11
<5 % UT (>95 % dip in
UT) for 5 sec
Power frequency
NOTE: UT is the a.c. mains voltage prior to the application of the test level
Electromagnetic Environment-Guidance
The device uses RF energy only for its internal function. Therefore, its
Group1
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment
Class B
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network.
Class A
Complies
Compliance
±6kV contact
±8kV air
±2kV for power
supply line
out line
1 kV line(s) to
line(s)
<5 % UT (>95 % dip
in UT) for 0,5 cycle
40 % UT (60 % dip in
UT) for 5 cycles
70 % UT (30 % dip in
UT) for 25 cycles
<5 % UT (>95 % dip
in UT) for 5 sec
18
Electromagnetic Environment-Guidance
are covered with synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment. If the user of this device requires
continued operation during power mains interruptions,
it is recommended that the device be powered from an
uninterruptible power supply or a battery.
characteristic of a typical location in a typical commer-
cial or hospital environment.