WARNINGS
• Filter solutions with a temperature limit of
45 °C (113 °F).
• Syringe filters are for manual use only; do not use
on automated systems .
• Not suitable for filtering high viscosity solutions
or blood .
• Do not filter solutions containing 5 mg or less of
active drug materials unless binding studies have
been performed .
• To ensure sterility, do not use this product if the
package is damaged .
• Use aseptic technique to maintain sterility removing
product from primary package .
• Do not use with syringes smaller than 10 mL because
pressures in excess of the maximum pressure rating
may be reached, potentially causing damage to the
filter and/or personal injury .
• Do not use for direct neuraxial applications which
may come in contact with cerebrospinal fluid .
• Do not use the same syringe filter to filter solutions
in both directions .
• Sudden loss of pressure could indicate failure
of the filter .
• Single use only; do not re-use or resterilize.
• Make sure to wet the filter membrane thoroughly
before injecting the solution; improperly wetted
filters can become airlocked .
• Not suitable for use with medical pressure
infusion equipment or gravity fed intravenous (IV)
administration sets .
• Discard appropriately after single use . See
"Disposal" below .
Directions for Use
1 . Wearing gloves using aseptic technique:
fill syringe with solution
to be filtered .
2 . Remove cover from package .
3 . Attach suitable syringe to filter inlet and
remove assembly from package .
4 . Hold syringe with filter pointing up and
top off by pushing a few drops through
to wet filter .
Y Excess solution may be hazardous
and should be disposed with care .
5 . Depress syringe plunger to filter
solution into suitable container or
compatible attachment on outlet .