USER'S MANUAL
HANDPIECE TROUBLE SHOOTING
PROBLEM
1. Compressor not ON.
2. Water Flow Control not ON (P4- E).
3. Water/CLS Bottle Empty.
No Water Flow in
4. Check for a 'kink' in the water line.
Handpiece
5. Check if 3 way syringe has Water.
6. Use of tap water, blocking water valves.
7. Low regulator pressure
1. Use of tap water - mineral or chemical
Mineral Accumulation
2. Verdigris (Green Slime) build up in water
1. Most brands of Chlorhex will deposit
Use of Chlorhexidine
2. Chlorine in tap water will cause a
EVOLUTION VETERINARY DENTAL UNITS
CAUSE
action blockage will usually take 3-4 years
to occur.
valves due to chemical reaction with tap
water.
out of solution and block valves, etc.
CLS is designed to be used in its diluted
strength without depositing. NOTE:
Do not dilute CLS beyond its label
recommendations or it could result in
deposits.
flocculation and block filters and valves
SOLUTION
1. Turn compressor ON and check that ON/
OFF tap is in the ON position.
2. Turn water control to ON.
3. Fill water/CLS Bottle(s).
4. Check lines and pressure.
5. If yes, lack of water must be in the valves
or the handpiece. The 3 way syringe is
the quickest way to check for water.
6. Valves will need replacing.
7. Check regulator is set at 70 PSI.
1. Pass a fine wire through the water intake
hole. Pass another fine wire through the
hole in the handpiece head. Pass the wire
as far as possible each way. 'Blow out' the
loosened mineral deposits.
2. Replace valves
1. Contact your local iM3 Official Service
Centre
2. Contact your local iM3 Official Service
Centre. Use only distilled water
56 � EN