USER'S MANUAL
LOW SPEED TROUBLE SHOOTING
PROBLEM
1. No or low air pressure.
2. Air line improperly sealed.
3. Inferior oil; too much or excess more
Handpiece won't run
4. Broken drive head. Note new handpieces
1. Plastic drive shaft broken off internally in
Can't fit
R&R prophy angle
2. Improper lubrication. Only fit plastic iM3
CARTRIDGE TOGGLE
Toggle switches become floppy
All iM3 dental units use a unique cartridge system for
the toggle switches. The toggles are used to pressurize
the water/CLS bottles . When the o-rings become worn
you simply push the cartridge in and turn a 1/4 turn anti
clockwise. The cartridge can then be removed from the
valve casing. To install a new toggle, locate the desired
position, push and rotate a 1/4 turn clockwise. New toggles
can be ordered from iM3 and installed by the user. Toggles
can be installed in the 12, 3, 6 and 9 o'clock positions. If
the toggle is in the wrong position simply remove and
reposition. The Dental unit should be turned off and
depressurised before replacement
ULTRASONIC SCALER HANDPIECE
For cleaning information read in full the sub
manual for the Scaler
EVOLUTION VETERINARY DENTAL UNITS
CAUSE
commonly too little oil or water in the
turbine.
require a "Run in time" this can be up to
20 hours of use
Chuck housing.
R&R prophy angles Code L7466.
SOLUTION
1. Check the air line to be sure the air is
turned on and the line is unobstructed.
2. Reconnect air line to handpiece making
sure the gasket is fully contracted.
3. See lubrication instructions on page 17.
4. Contact your local iM3 official Service
Centre.
If the motor stops spinning place a fine
screw driver down the nose and rotate
shaft.
1. Contact your local official iM3 Service
Centre.
2. See lubrication instructions
Fig.48 - Toggle switch for Water and CLS
57 � EN