Dräger DAP 5000 Instrucciones De Uso página 15

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 16
6. Pay attention to the preferred position between the inlet
connection and the 90° angle inside the water container.
The inlet connection and angle inside the water container
must be positioned at an angle of 180° to one another. If
necessary, vary the number of flat gaskets.
6.3.2
Replacing the profile gaskets
1. Loosen both tensioning straps.
2. Remove the lid. Pay attention to the soda lime's handling
instructions.
3. Remove the lime container and carefully place it on the
floor.
4. Remove the old profile gasket from the upper edge of the
lime container and the water container.
5. Insert the new profile gasket into the upper edge of the lime
container and the water container. Cut the gasket to size so
that the two ends create a contact seal. We recommend
only cutting the gasket to size during insertion.
6.3.3
Replacing the return valve's valve disc
1. Prepare the DAP 5000 (see chapter 6.2.1).
2. Remove the old valve disc from the valve seat by pulling.
3. Insert the new valve disc on the valve seat.
6.3.4
Replacing the fill level indicator
1. Prepare the DAP 5000 (see chapter 6.2.1).
2. Loosen the nuts inside the water container and remove the
old fill level indicator.
3. Fit the new fill level indicator and tighten the nuts inside the
container with a tightening torque of 6 Nm.
6.3.5
Assembling the feet
1. Prepare the DAP 5000 (see chapter 6.2.1).
2. Clean the adhesive surfaces on the base of the water
container of contamination.
3. Firmly push the adhesive side of the foot onto the base of
the water container.
7
Transport and storage
Completely drain and dry the water container so that no
bacteria can grow (see chapter 6.2.1). Completely drain and
clean the lime container (see chapter 6.2.2). Close the inlet
and outlet connector to protect against contamination.
Observe the temperature range (see chapter 9). Protect the
product against direct sunlight.
Store the product in the supplied original packaging during
extended periods of storage or transport.
8
Disposal
The product, the water and the soda lime must be disposed of
in accordance with the national waste disposal regulations or
by an appropriate waste disposal company.
9
Technical data
Weight (not including the soda
lime and water)
Dimensions (W x D x H)
Connection, inlet connector
Connection, outlet connector
Max. CO
concentration of the
2
ambient air
CO
outlet concentration
2
Medium
Air flow
Permitted negative pressure
Typical pressure drop on the DAP
5000
Maximum daily operating time
Rel. humidity of the intake air
Operating temperature
Storage temperature
10
Order list
Designation
Soda lime canister Dive Sorb S - 18 kg
Filter mat (10 pieces)
Spare part set (fill level indicator, flat
gaskets, profile gaskets, valve disc, feet)
Sterilisation tablets (40 pieces)
Hose connection set (2x hose nipple,
diameter 32-38, 40 and 50 mm)
Intake hose, inlet side
60 mm/60 mm (2 m length)
Intake hose, outlet side
60 mm/60 mm, for open systems
60 mm/40 mm, for open systems
60 mm/60 mm, for closed systems
Filling funnel (suitable for the soda lime
canister Dive Sorb S - 18 kg)
Transport and storage
en
approx. 23 kg
455 x 800 x 330 mm
Diameter 63 mm
Diameter
1000 ppm
≤500 ppm
Air
0 to 700 L/min
0 - 20 mbar
approx. 8 mbar at 700 L/min
5 hours
10 to 100 %
+5 °C to +45 °C
-10 °C to +50 °C
Ordering no.
67 37 111
69 13 819
69 13 818
69 13 282
69 15 525
65 98 080
65 98 080
65 98 082
65 98 085
69 22 767