Declaration Requirements – UK SI 2016/1107 (as
amended by SI 2019/696)
This equipment is designed and manufactured to meet the
essential health and safety requirements not covered by
UK-Type Examination Certificate CSAE 21UKEX2356X
when installed as detailed above.
Pressure Safety
This equipment meets the requirements of the European
Pressure Equipment Directive 2014/68/EU and
UL 61010-1 for pressure safety.
Return Goods/Material Procedure
If the unit requires calibration or is unserviceable, return it
to the nearest Druck Service Centre listed at:
https://druck.com/service.
Druck is an active participant in the UK and EU Waste
Electrical and Electronic Equipment (WEEE) take-back
initiative (UK SI 2013/3113, EU directive 2012/19/EU).
The equipment that you bought has required the
extraction and use of natural resources for its production.
It may contain hazardous substances that could impact
health and the environment.
In order to avoid the dissemination of those substances
in our environment and to diminish the pressure on the
natural resources, we encourage you to use the
appropriate take-back systems. Those systems will
reuse or recycle most of the materials of your end life
equipment in a sound way. The crossed-out wheeled bin
symbol invites you to use those systems.
If you need more information on the collection, reuse,
and recycling systems, please contact your local or
regional waste administration.
Please visit the link below for take-back instructions and
more information about this initiative.
https://druck.com/weee
Contact the Service Department to obtain a Return
Goods/Material Authorization (RGA or RMA). Provide the
following information for a RGA or RMA:
Product (e.g. DPI705E-IS)
Serial number.
Details of defect/work to be undertaken.
Calibration traceability requirements.
Operating conditions.
Quick Start
Key
Description
Switches instrument on or off.
Filter on or off.
Tare on or off (changes the display reading to zero).
Changes units (1 of 16 pressure units, °C or °F).
View maximum, minimum or perform leak test (Max/Min
is reset at switch on).
Back light. Short press illuminates for 5 seconds. Long
press toggles between permanent on and off.
Time Out
If a key is not pressed within 10 minutes, the instrument
times out and switches off.
Leak Test
To perform a leak test, press the
timer symbol
shows on the display with a leak test
duration of 60 seconds. Press the
duration to 120 or 300 seconds. To start the leak test, press
the
key again. The instrument counts down to zero,
then displays the leak rate. Press the
during the leak test to quit and return to normal
measurement.
Zero
A zero should be performed on gauge and differential
instruments before measuring pressure. To perform a zero,
open all pressure ports to atmospheric pressure. Press the
and
keys together. The display briefly shows
and the instrument calculates a new zero.
English–DPI705E / DPI705E-IS Safety Manual | 3
key three times. The
key to increase the
key at any time
Copyright 2019 Baker Hughes Company.