Disinfection and decontamination:
Exterior casing and mechanical surfaces must be wiped with a clean cloth, dampened with a disinfectant
Soft plastic parts and rubber surfaces can be cleaned by gently wiping a clean cloth, dampened with a
disinfectant. Discoloration can occur if alcohol is used
The front lens of eyepieces and objectives are sensitive to chemicals. We recommend not to use aggressive
disinfectants but to use lens paper or a soft fibre-free tissue, damped in cleaning solution. Cotton swabs may
also be used. We recommend you use personal eyepieces without eyeshades in order to minimize risk
Never immerse or dip the eyepiece or objective into a disinfectant liquid! This will damage the component
Never use abrasive compounds or cleaners that may damage and scratch optical coatings
Properly clean and disinfect all possible contaminated surfaces of the microscope or contaminated accessories
before storing for future use. Disinfection procedures must be effective and appropriate
Leave the disinfectant on the surface for the required exposure time, as specified by the manufacturer. If the
disinfectant evaporates before the full exposure time, reapply disinfectant on the surface
For disinfection against bacteria, use a 70% aqueous solution of isopropanol (isopropyl alcohol) and apply for at
least 30 seconds. Against viruses, we recommend to refer to specific alcohol or non-alcohol based disinfection
products for laboratories
Before returning a microscope for repair or maintenance through a Euromex dealer, an RMA (return authorization
form) together with a decontamination statement must be filled in! This document - available from Euromex for any
reseller- must be shipped together with the microscope at all times
Reference documents:
World Health Organisation:
https://www.who.int/ihr/publications/biosafety-video-series/en/
Robert Koch Institut:
https://link.springer.com/content/pdf/10.1007/s00103-013-1863-6.pdf
US Centre for Disease Control and prevention
https://www.cdc.gov/infectioncontrol/guidelines/disinfection/index.html
Handle with care
This product is a high quality optical instrument. Delicate handling is required
Avoid subjecting it to sudden shocks and impacts
Impacts, even small ones, can affect the precision of the instrument
Handling the LED
Note: Always disconnect the power cord from your microscope before handling the LED bulb and power unit and
allow the system to cool down approximately 35 minutes to avoid burns
Never touch the LED with your bare hands
Dirt or fingerprints will reduce the life span and can result in uneven illumination, lowering the optical
performance
Use only original Euromex replacement LEDs
The use of other products may cause malfunctions and will void warranty
During use of the microscope the power unit will get hot; never touch it while in operation and allow the system
to cool down approximately 35 minutes to avoid burns
Dirt on the lenses
Dirt on or inside the optical components, such as eyepieces, lenses, etc., affects the image quality of your system
negatively
Always try to prevent your microscope from getting dirty by using the dust cover, prevent leaving fingerprints
on the lenses and clean the outer surface of the lens regularly
Cleaning optical components is a delicate matter. Please, read the cleaning instructions further on in this
manual
4