Euromex EduBlue BlueLine Serie Manual Del Usuario página 9

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 35
2. The correct eye point
The eye point is the distance from the eyepiece to the user's pupil. To obtain the correct eye point, move the eyes
towards the eyepieces until a sharp image is reached at a full field of view
3. Adjusting the diopter
In order to obtain the right interpupillary setting, one should go through the below steps
Turn to the highest magnification
Turn the diopter adjustment ring of the left eyepiece tube until the scale shows the same reading as on the
indicator
Close the right eye and focus the left tube by means of the coarse adjustment knobs
Close the left eye and focus the right tube with the diopter adjustment ring
This procedure should be followed by each individual user. When more users are working on the same
stereomicroscope it is recommended to remember your own diopter setting for a quick set up during new microscopy
sessions
Adjusting the focusing tension
When the system lowers itself by its own weight, the tension of the focusing
knob (D) should be adjusted. This is done by turning the two focusing knobs
in opposite (clockwise or counterclockwise) directions (see image on the
right)
Illumination & brightness adjustment
All models have a transmitted (diascopic) and incident LED illumination. With
the potentiometers (F) both illuminations can be switched ON and OFF and
adjusted in strength
The potentiometer on the right is used to adjust the bottom illumination,
the one on the left is used to adjust the top illumination
The illumination has the following specifications:
LED:
External power supply:
1W LED
Primary AC 100 - 240 Volt 50/60Hz.
9