SELECTING THE STERILIZATION PROGRAMMES
The preheating time from cold is 10 minutes.
By selecting the submenu "Programs", the sterilization cycles that the autoclave is able
to carry out can be accessed:
1 - 134°C Standard type B cycle
2 - 134°C Prion type B cycle
3 - 121°C Standard type B cycle
4 - 134°C B Flash type B cycle*
The selection indicator is positioned on "134°C Standard B" cycle.
To scroll the cycles and move the selection indicator, press the (+) or (-) keys, then confi rm
with the
Press the "\\" key to cancel the operation.
STANDARD CYCLES
- Cycle at 134°C
- Cycle at 134°C B Flash* : 3,5 minutes of sterilization plus 8 minutes of drying
- Cycle at 121°C
*
Maximum load: 0,5 Kg of packaged material – 0,2 Kg of porous material..
NOTE: For each available sterilization cycle must be done by manufacturer the Works Test as specifi ed by EN
13060 (par. 7.3.1). The available cycles for that the manufacturer hasn't done the Works Test must be performed by
Installation Tests following the EN 13060 par. 7.4.
PRION CYCLE
- Cycle at 134°C
"LAST ADVANCED CYCLE" FUNCTION
The "Last Advanced Cycle" function is enabled for all programs.
30 minutes after the end of sterilization cycle, the heating elements switch off and a se-
cond drying cycle starts automatically when the temperature inside the chamber reaches
60°C.
MODIFYING STERILIZATION TIMES
The sterilization times can be increased; this operation requires a specialized technician.
30
The packaged items must be placed on the trays with the transparent
part of the packet facing downwards.
key.
:
: 16 minutes of sterilization plus 14 minutes of drying
: 18 minutes of sterilization plus 14 minutes of drying
4 minutes of sterilization plus 14 minutes of drying