Guidance and manufacturer´s declaration
– electromagnetic emission –
for all EQUIPMENT AND SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic emission
The model TT-CL012 is intended for use in the electromagnetic environment
specified below. The customer or the user of the model TT-CL012 should assure
that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic
Class A
emissions
IEC 61000-3-2
Voltage fluctua-
Applicable
tions /flicker
emissions
IEC 61000-3-3
13
Electromagnetic environment –
guidance
The Model TT-CL012 uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
The TT-CL012 is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Guidance and manufacturer's declaration –
electromagnetic immunity –
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The Model TT-CL012 are intended for use in the electromagnetic environment
specified below.The customer or the user of the Model TT-CL012 should assure
that it is used in such an environment.
Immunity test
IEC 60601
Compliance level
test level
Electrostatic
± 8 kV contact
± 8 kV contact
discharge (ESD)
±2 kV, ±4 kV,
±2 kV, ±4 kV, ±8 kV,
IEC 61000-4-2
±8 kV, ±15 kV
±15 kV air
air
Electrostatic
± 2 kV
± 2 kV for power
transient / burst
for power
supply lines
IEC 61000-4-4
supply lines
100 kHz repetition
100 kHz repe-
frequency
tition frequen-
cy ± 1 kV for
input/output
lines
EN
Electromagnetic
environment -
guidance
Floors should be wood,
concrete or
ceramic tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30 %.
Floors should be wood,
concrete or
ceramic tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30 %.
14