HIAC 9703+ Liquid Particle Counter
Specifications
Specifications
Specifications are subject to change without notice.
Specification
Dimensions (W x D x H)
Weight
Power requirement
Viscosity limits
Coincidence loss limit
Operating temperature
Operating humidity
Maximum altitude
Sample bottle clearance
Sample temperature
Flow rate
Flow rate accuracy
Volume accuracy
Tare volumes for probes
Sensor concentration limits
(particle per ml)
Wetted materials
Number of channels
Number of bins
Certification
a. Under ideal conditions, inaccuracy in flow rate is ≤1% of syringe size.
2 / 22
Details
343 x 337 x 482 mm (13.5 x 13.3 x 19.0 in.)
10.7 kg (23.6 lb)
External Class III power adapter: 100–240 VAC, 50–60 Hz, 2.5 A input; 24 VDC,
3.75 A output
Instrument: 24 VDC, 75 W maximum
< 50 cp
7,500
5 °C to 40 °C (41 °F to 104 °F)
0 to 80% relative humidity for temperatures up to 31 °C decreasing linearly to
50% RH at 40 °C
2,000 m (6,560 ft)
153 mm (6.02 in.)
5 °C to 40 °C (41 °F to 104 °F)
10 to 100 mL/min (determined by the sensor calibration flow rate)
> 95%
a
> 95% of syringe size
0.091 mL for 8.1 cm (3.2 in.)/1.2 mm (0.047 in.) ID
0.172 mL for 15.4 cm (6.1 in.)/1.2 mm (0.047 in.) ID
1.57 mL for 14 cm (5.5 in.)/large bore
HRLD150: 18,000
HRLD150JA: 18,000
HRLD400: 10,000
HRLD600JS: 6,000
MC05: 9,000
®
PTFE Teflon
, TFE Teflon, PFA Teflon, Kel-F, 316 Stainless steel, Glass, Viton
4 (2 for light scatter, 2 for light extinction)
65,536 (16,384 bins per channel)
CE
®
DOC026.98.80097, Edition 7