7.
OPERATION
CAUTION
This product emits potentially hazardous optical
radiation. Do not stare directly into the light cone. Eye
irritation may occur.
The radiation emitted by this product complies with the
exposure limits for reducing the risk of photobiological
hazards based on IEC 62471: RG 2 (medium risk).
1
2
1
2
Switch the luminaire on or off using the button 1.
The luminaire can be dimmed using the +/- buttons 2
The light color can be adjusted using button 3
The focus can be adjusted using button 4
The endoscope mode can be adjusted using button 5
Before each use, perform a function test: all LEDs in the
light cone must come on.
Triango 100 F
DANGER
Electric shocks are life-threatening.
Do not plug in any damaged power cables.
If there are any signs of damage to the power cable,
immediately replace it with a new one.
Connection voltage and frequency must match data
on the type plate.
Only connect the luminaire to the supply network with
a protective earth conductor (PE).
DANGER
In the event of a change of location
Put the terminal into its lowest position.
Unlock the castors.
Take care not to run over anything, including the
connection cable.
Hold firmly during transport within the hospital.
Take particular care on slopes, thresholds, bumps and
other obstacles.
CAUTION
Triango 100 F
Wind the power cable onto the cable holder when not
in operation
28
3
4
5
Plug in cable
Connect cable to the network
Lock the castors
8.
CLEANING AND DISINFECTION
DANGER
Death hazard from electric shock
Before disinfection cleaning. pull the plug out of the
socket and secure the device against being turned
back on.
CAUTION
Material damage due to incorrect cleaning
For cleaning, use only products that do not impair
operation of the luminaire.
For cleaning, do not use solvent-based, chlorine-
based or abrasive detergents, because they can crack
plastic parts and cause other damage.
The cleaning agents must be approved for use on
plastics such as PC, PMMA, PA and ABS.
Concentrated disinfectant can damage the luminaire.
For concentration and application times, check the
information provided with the product used.
The wrong cloth may cause scratches.
RECOMMENDED DISINFECTANTS
Bacillol 30 Foam
Dismozon Plus
Kohrsolin Extra
Lysoformin
Mediclean
Meliseptol Foam Pure
Microbac Tissues
Mikrozid Sensitive Liquid
Mikrozid AF Liquid
Sagrotan
Terralin Protect
Virex Tb
CaviCide 1
Germicidal Bleach
Hexaquart XL