Descargar Imprimir esta página

DJO Compex edge Manual Del Usuario página 40

Publicidad

Idiomas disponibles

Idiomas disponibles

SPECIFICATIONS
TABLES 2 & 3: RECOMMENDATIONS AND DECLARATION
ELECTROMAGNETIC IMMUNITY
The Compex Edge is designed for use in the electromagnetic environment stipulated
below. The customer or the user of the Compex Edge must ensure that it is used in
this recommended environment.
Immunity
Test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/
burst
IEC 61000-4-4
Surge
(1)
IEC 61000-4-5
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power
frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
BY THE MANUFACTURER -
IEC 60601
Test Level
±6 kV at the
contact
±8 kV in air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV differential
mode
±2 kV common mode
<5% U
(>95% dip in
T
U
) for 0.5 cycle
T
40% U
(60% dip in
T
U
) for 5 cycles
T
70% U
(30% dip in
T
U
) for 25 cycles
T
<5% U
(>95% dip in
T
U
) for 5 sec
T
3 A/m
Compliance
Level
±6 kV at the
Floors must be wood,
contact
concrete, or ceramic tile.
If floors are covered with
±8 kV in air
synthetic material, the
relative humidity should be
at a minimum of 30%.
Not Applicable
The quality of the power
supply should be that of
System battery-
a typical commercial or
powered
hospital environment.
Not Applicable
The quality of the power
supply should be that of
System battery-
a typical commercial or
powered
hospital environment.
Not Applicable
The quality of the power
supply should be that of
System battery-
a typical commercial or
powered
hospital environment. If the
Compex Edge user requires
continuous operation
during mains power cuts,
we recommend that the
Compex Edge is powered by
a UPS or a battery.
Magnetic fields at the mains
frequency should be at the
level of a representative
site located in a typical
commercial or hospital
environment.
35
Electromagnetic
Environment -
Guidance

Publicidad

loading