During the initial power-on testing or during operation, it is possible the scale may show an error
message. Specific error message info is described below.
If an error message is shown, repeat the procedure that caused the message (turning the
balance on, calibrating or other functions). If error messages are displayed, turn scale off, wait
3-5 minutes and then turn scale on again. If the error message is still shown, contact Uline
Customer Service at 1-800-295-5510 for further support.
ERROR CODE
Err4
Err5
Err6
Err7
Err8
Err9
PAGE 3 OF 9
ERROR CODES
DESCRIPTION
Initial zero is greater than
allowed (4% of maximum
capacity) when power
is turned on or when the
ZERO/ENTER key is pressed.
Keyboard error.
A/D count is not correct
when turning the scale on.
Can't set current weight as
100%.
Calibrate weight error.
Unstable, can't return
to zero. (This message
only displays as scale is
powering on.)
Weight on the pan when turning the scale on.
Excessive weight on the pan when zeroing the
scale.
Improper calibration of the scale.
Damaged load cell.
Damaged electronics.
Improper operation of the scale.
Platform not installed.
Load cell damaged.
Electronics damaged.
This message only shows when you press
in zero point.
Incorrect weight used.
Verify scale is on a stable surface.
If scale was outside of operating temperature
range, allow ample time for unit temperature
to return to acceptable range.
(See Specifications section.)
1-800-295-5510
uline.com
CAUSES
%
key