Further basic functions and settings

NOTE
32
Further basic functions and set-
tings

Internal flushing

To ensure that the analysed sample represents the current value, the
sampling line must be sufficiently flushed.
Duration of the internal flush time
The duration of the internal flush time is preset (10 s before and
after a measurement) and cannot be influenced by the operator.
The quantity of flush water for internal flushing depends on the pres-
sure:
Pressure
Max. water quantity
per analysis
1 bar
2 bar
3 bar

External flushing

If a very long (approx. 9.8 – 32.8 ft (3 – 10 metres)) sampling line or a
line with a large diameter is used, install an external flush valve up-
®
stream of Testomat
808. Connect the external flush valve to the
"Relay 1" outlet.
Flush process – internal/external in manual mode
Proceed as follows to additionally flush the instrument:
At first switch the instrument into standby mode for flushing. Short
press key 3 (pause) to achieve this.
If a measurement is running, wait until the measurement has been
completed.
Internal flush process:
Short press key 2 (Flush) to open the internal flush valve.
The valve opens and the measuring chamber is flushed.
The LED above key 2 (Flush) flashes as long as the internal flush
valve is activated.
Short press key 2 (Flush) again to terminate the flush process.
80 ml
120 ml
160 ml