Handling measuring
68 •
results
Do not expose the device to any impacts or
vibrations.
Perform function controls regularly as de-
scribed in the relevant section in this docu-
ment. Do not operate the device if it is
damaged or not working properly.
Ensure that there is no heat source in the im-
mediate vicinity. Do not expose to direct sun-
light. The excessive temperature could
damage the electronics.
Avoid rapid temperature fluctuations. When
the device is transported so that a tempera-
ture difference of more than 20 °C occurs, it
must stay turned off for at least 2 hours be-
fore it can be turned on again. Otherwise,
condensation water will form which can dam-
age the electronics.
Use the device only in the ambient conditions
outlined in "Intended use".
Store the device only in the storage condi-
tions outlined in "Intended use".
Use only chlorine and alcohol-free disinfec-
tants which are explicitly suitable for acrylic
sheet and other sensitive surfaces (active in-
gredient: quaternary ammonium com-
pounds, for example).
Do not use aggressive or abrasive cleaning
agents.
Do not use organic solvents (e.g. white spirit
or petroleum spirit).
WARNING!
Patient hazard
This device is no diagnostic device. It simply
assists the treating physician in establishing a
diagnosis.
In order to make a precise diagnosis and ini-
tiate therapeutic measures, besides determi-
nation of the weight, further targeted
examinations must be set up by the physi-
cian, and their results must be considered.
The responsibility for diagnosis and treatment
lies with the treating physician.