General Description; Meter - Alnor Velometer 6000 Serie Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

A Range
Scales
C Zero
Adjustment
B Sensing
Screw
Ports

GENERAL DESCRIPTION

The Alnor Velometer is a direct reading instrument for measuring air
velocities. It is designed to measure velocities inside heating and
ventilating ducts, or in open areas such as at fume hoods, grilles,
diffusers, slots on ventilated plating tanks, and so forth. It may also be
used for measuring duct static pressures.
The Velometer set consists of the meter, velocity and static pressure
Range Selectors, measuring probes and connecting hoses.
The Velometer is basically an air flow meter reading in feet per minute
(or meters per second).
The Velometer is built with a taut band meter movement; it is carefully
balanced so that you may hold the meter in any normal position and still
get an accurate reading. All Velometers are uniformly calibrated to
factory standards: optimum accuracy is attained when velometers are
used with accessories of the same set, and with the meter in the
horizontal position (scale window facing up).
Before using the Velometer, study the operational instructions and
procedures to ensure proper use of the instrument for accurate results.

Meter

The meter receives air through its sensing ports. A force against the vane
causes a deflection of the needle which yields an indicated reading.
On the face of the meter there is a series of range scales (A) which
enables you to read accurately the measured value. Since there are a
number of ranges, it is important that the scale read on the meter face
corresponds to the velocity or static pressure range that is set on the
Range Selector.
On the back side of the meter there are two sensing ports (hose
connections) (B) which are for interconnecting to one of the two Range
Selectors with hoses. One port is marked positive (+) and the other
negative (–).
Sensing port caps are provided and you should place them over the
sensing ports when the meter is not in use.
Check to see that the pointer is at zero each time the meter is put into
service. The zero adjustment screw (C) enables you to make an
adjustment. With the caps still on the ports, use a screwdriver to bring
the pointer to zero. It should continue to read zero whether the meter is
lying flat, standing upright or is at any position in between, including the
position where one end of the scale is lower than the other.
3

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido