CAUTION
• Clean and lubricate the handpiece prior to autoclaving. If blood remains on the internal surface it can become clotted and cause product
failure.
• Do not autoclave the product with other instruments even when it is in a pouch. This is to prevent possible discoloration and damage to the
product from chemical residue on other instruments.
• Do not heat or cool the product too quickly. Rapid change in temperature could cause damage to the product.
• To avoid product failure, do not use a sterilizer that exceeds a cycle temperature of 138°C, including the dry cycle. In some sterilizers, the
chamber temperature may exceed 138°C. Contact the sterilizer manufacturer for detailed information about cycle temperatures.
• Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. The air should be free from dust, salt and
sulphur.
• Do not touch the product immediately after autoclaving as it will be very hot and must remain in a sterile condition.
• Autoclave sterilization is recommended for the product. The validity of other sterilization methods is not confirmed.
NOTICE
• NSK recommends Class B sterilizers as stated in EN 13060.