NSK S-MAX M M15L Manual Del Usuario página 3

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 66
WARNING
• Immediately after a treatment, the handpiece should be cleaned, lubricated and sterilized. Failure to properly maintain the
handpiece may lead to overheating, causing burn injuries or product failure. (Refer to "7. Post-use Maintenance")
• If blood infiltrates inside a handpiece, an automatic handpiece cleaning and lubrication system may not totally clean the
internal handpiece components and may lead to blood coagulation inside the handpiece. Coagulated blood may cause
handpiece failure and overheating, causing burn injuries. To avoid the risk, NSK recommends to immediately lubricate those
handpieces with PANA SPRAY Plus / PANA SPRAY.
• Depressing the Push Button while the handpiece is in rotation may lead to overheating, causing burn injuries or product
failure. Avoid the push button to contact with any oral tissue.
• Supply coolant water and coolant air while using the handpiece. No supplying the coolant water and coolant air may lead to
overheating, causing burn injuries or product failure.
• Do not use the handpiece for prophylaxis purposes. Polishing paste could enter the handpiece, causing the failure of the
Push Button or cause burn injuries due to the overheating of the handpiece head.
• Keep any debris or other foreign materials away from inside the gear or the handpiece. Foreign materials remaining inside
may lead to overheating, causing burn injuries or other accidents.
2

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

S-max m m25lS-max m m95lS-max m m15S-max m m25S-max m m95

Tabla de contenido