NSK automatic handpiece cleaning and lubrication system
When using NSK automatic handpiece cleaning and lubrication system, refer to the system instructions.
WARNING
• If blood infiltrates inside a handpiece, an automatic handpiece cleaning and lubrication system may not totally
clean the internal handpiece components and may lead to blood coagulation inside the handpiece. Coagulated
blood may cause handpiece failure and overheating, causing burn injuries. To avoid the risk, NSK recommends
to immediately lubricate those handpieces with PANA SPRAY Plus / PANA SPRAY.
7-6 Sterilization
Sterilize the product by autoclave sterilization. After each patient sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 20min. at 121°C, or 15min. at 132°C, or 3min. at 134°C.
3) The handpiece should remain in the autoclave pouch until required for use.
15