SUGGESTED DECONTAMINATION
PROCEDURES
Existing alkaline solutions for the pre-disinfection of certain medical devices are NOT RE-
COMMENDED for pre-disinfecting our cameras.
It is essential for the parts coming into contact with the disinfectant to be thoroughly rinsed.
Use soft non-woven cloths for wiping the lenses dry so as not to scratch them.
The procedures described in this section are given as guidance, and cannot under any
circumstances be substituted for official recommendations or directives.
The connector on the camera equipped with its sealing cap can be immersed. Traces of
water are visible just as the connector seal is removed (runoff and condensation). It is therefore
essential to dry the inside of the cap well before reusing it.
Disinfecting the lens surfaces
A dirty lens surface can interfere with observation. The lens surface must not have any marks or
smudges. To avoid scratching the lens surface, never use abrasive cloths or sponges to clean it.
To remove dust, dirt and other impurities not originating from the patient, wipe clean with a soft,
lint-free cloth moistened with ethanol or 70% isopropyl alcohol. When cleaning, wear gloves
for protection from chemical products.
Ensure that the equipment is completely dry before use.
Use soft non-woven cloths to clean the lenses so as to avoid scratching them.