Fault
Cause
code
E00
Supply voltage is below 9 V
or above 30 V
E01
Hardware failure
E08
System failure
E09
Communication failure
between remote access
box and gas transmitter
E10
Display failure
(software on the remote
access box not compatible
with the gas transmitter
software)
E11
SIL lock of the gas
transmitter has been
configured to "on" when the
calibration or configuration
menu on the remote access
box is active
E12
Configuration failure
If the faults occurred cannot be resolved using the remedy measures
described or if other not described faults occur, the remote access box
must be checked by the Dräger service department and where
applicable repaired.
Remedy
Connect transmitter to
proper voltage
Contact Dräger service
department: Have the
converter module replaced
Contact Dräger service
department: Have the
converter module replaced
Check the electrical
installation and status of
the gas transmitter
Contact Dräger service
department: Have the
software on the remote
access box updated
Configure the SIL lock of
the gas transmitter to "off"
(see technical manual for
Dräger PIR 7000 / Dräger
PIR 7200)
Check the configuration of
the gas transmitter
Description of design
The remote access box consists of a housing, a flameproof converter
module (with integrated electronics and software), a cable gland and
a blind plug.
The housing made of conductive plastic prevents electrostatic charge-
up and is resistant to solvents.
The connection of the remote access box to the gas transmitter is
carried out via a three-wire line, with communication between the
remote access box and the gas transmitter is carried out via the serial
interface (ICOM, white connection lead).
Technical data
Signal transmission to the
Serial interface (ICOM,
gas transmitter
white connection lead)
Supply voltage
9 to 30 V DC, nominal 24 V DC
Power consumption
≤ 1.0 W
Electrical connection
Cable diameter 7 to 12 mm,
Cross-section 0.5 to 1.5 mm
Dimensions
approx. 110 x 150 x 60 mm (W x H x D)
Weight
approx. 250 g
IP rating
IP 66
CE mark
Equipment and protective systems for
intended use in potentially explosive
atmospheres (Directive 2014/34/EU)
2