Medisana MTR Instrucciones De Manejo página 34

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 65
6 Miscellaneous
GB
6.2
Care and
maintenance
6.3
Disposal
26
"Er 0": Pressure system is unstable before measurement. Fit cuff
correctly and tighten and repeat the measurement according to the
instructions for use.
"Er 1": Fail to detect systolic pressure. Please repeat the measurement
according to the instructions for use.
"Er 2": Fail to detect diastolic pressure. Please repeat the measurement
according to the instructions for use.
"Er 3": Pneumatic system blocked or cuff is too tight during inflation.
Fit cuff correctly and tighten and repeat the measurement according to
the instructions for use.
"Er 4": Pneumatic system leakage or cuff is too loose during inflation.
Fit cuff correctly and tighten and repeat the measurement according to
the instructions for use.
"Er 5": The pressure in the cuff is too high (over 300 mmHg) or the
pulse (over 200 beats per minute) is too high. Relax for 5 minutes and
repeat the measurement.
"Er 6": Inflation takes too long. Fit cuff correctly and tighten and repeat
the measurement.
"Er 7": There is an electronical problem with the unit. Remove the
batteries and disconnect the adaptor cable. Wait for 5 minutes and
re-connect the power supply. Repeat the measurement according to
the instructions for use.
"Er 8": Device parameter checking error. Relax for 5 minutes and
repeat the measurement according to the instructions for use.
"Er A": Pressure sensor parameter error. Relax for 5 minutes and
repeat the measurement according to the instructions for use.
Important: Please contact the service centre if the error message
appears again when the measurement is repeated. Keep still and do
not move your arm or the cuff during measurement. The unit is not
suitable for people with an extremely irregular heartbeat.
• Remove the batteries before cleaning the unit.
• Clean the unit using a soft cloth lightly moistened with a mild soapy
solution.
• Never use abrasive cleaning agents or alcohol.
• Do not expose the unit to direct sunlight and protect it from dust
and moisture.
This product must not be disposed together with the domestic
waste. All users are obliged to hand in all electrical or electronic
devices, regardless of whether or not they contain toxic subs-
tances, at a municipal or commercial collection point so that
they can be disposed of in an environmentally acceptable
manner. Please remove the battery before disposing of
the equipment. Do not dispose of old batteries with your
household waste, but dispose of them at a battery collection
station at a recycling site or in a shop. Consult your municipal
authority or your dealer for information about disposal.

Publicidad

Tabla de contenido
loading

Tabla de contenido