Medical Devices; Loading Mesh Trays - Miele professional A 103 Instrucciones De Manejo

Cesto superior
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 39
en - Areas of application
Loading mesh
trays
Hollow instruments
34

Medical devices

Hinged instruments must be opened to minimise the risk of surfa‐
ces not being accessed during reprocessing.
To prevent the risk of light instruments and other small items being
moved about during reprocessing, place them in closable inserts
such as the E 473/1 or secure them in silicone holders such as the
E 476, E 477 and E 479.
Up to 6 hollow instruments can be reprocessed in the A 202 mobile
unit using a Luer lock connector.
To do so replace the blind stopper on the water pipe with a silicone
hose with Luer lock adapter, e.g. the E 448.
Connect instruments with the silicone hoses and place the instru‐
ments in a mesh tray on the upper level of the mobile unit. Secure
light instruments with silicone holders such as the E 476, E 477 and
E 479.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

A 202

Tabla de contenido