To avoid false readings, which could lead to possible electric shock or personal injury, replace
the battery as soon as the battery indicator
To replace the battery:
1. Disconnect the connection between the testing leads and the circuit under test, and remove
the testing leads away from the input terminals of the Meter.
2. Press the Meter power off.
3. Remove the screw from the battery compartment, and take out the battery door.
4. Remove the battery from the battery compartment.
5. Replace the battery with a new 9V battery (NEDA 1604 or 6F22 or 006P).
6. Rejoin the battery door and the battery compartment, and install de screw.
To avoid electrical shock or arc blast, or personal injury or damage to the Meter, use specified
fuses ONLY in accordance with the following procedure.
To replace the Meter's fuse:
1. Disconnect the connection between the testing leads and the circuit under test, and remove
the testing leads away from the input terminals of the Meter.
2. Press the Meter power off.
3. Remove the holster from the Meter.
4. Remove the screws from battery compartment, and then take out the battery door.
5. Remove the screw inside the battery compartment and also the other two screws from the
case bottom, and then separate the case bottom from the case top.
6. Remove the fuse by gently prying one end loose, and then take out the fuse from its bracket.
7. Install ONLY replacement fuses with the identical type and specification as follows and make
sure the fuse is fixed firmly in the bracket. 0.315A. 250V fast type, Ø5 x 20mm.
8. Rejoin the case bottom and case top, and reinstall the screws.
9. Rejoin the holster and the Meter.
Replacement of the fuses is seldom required. Burning of a fuse always results from improper
operation.