you continue to press the MEM button, the most recent
individual measured values with date and time are dis-
played in turn.
• You can delete the memory by pressing and holding the
MEM button for 3 seconds. All the values in the current
user memory are deleted after three beeps are output.
• To switch off the device, press the MEM button again or
the START/STOP button
vation switch to OFF.
• If you forget to switch off the device, it will switch off
automatically after 2 minutes.
8. Transferring measurements
Connect the blood pressure monitor to your PC using the
USB cable.
No data transfer may be launched whilst performing a
measurement.
PC is shown on the display. Begin the data
transfer in the "HealthManager" PC software.
During the data transfer, an animation is shown
on the display. A successful data transfer is
displayed as in figure 1.
or push the touch screen acti-
figure 1
If the data transfer is unsuccessful, an error
message appears as in figure 2. In this case,
interrupt the PC connection and start the data
transfer again.
After 30 seconds of not being in use or if com-
munication with the PC is interrupted, the blood
pressure monitor switches itself off automatically.
9. Error messages/trouble-shooting
In case of faults, the
Error messages can occur when
• systolic or diastolic pressure could not be measured (
or
appears on the display),
• systolic or diastolic pressure was outside the measure-
ment range ( or Lo appears on the display),
• the cuff is fastened too tightly or loosely (
appears on the display),
• the pump pressure is higher than 300 mmHg (
appears on the display),
• pumping up takes longer than 160 seconds (
on the display),
• there is a system or device error (
appears on the display),
• the batteries are almost empty
• the data could not be sent to the PC (
the display).