Beurer BM 93 Cardio Instrucciones De Uso página 53

Tensiómetro con función de ecg
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 87
Problem /
Error  Message
The ECG measurement
shows EE on the display.
The device will not switch
on.
The cuff does not inflate.
The blood pressure
measurements are very
high/low.
Possible cause
- Contact pressure on the
skin is too weak.
- The measurement was
interrupted.
- The sound level was too
high during the measure-
ment.
The batteries are almost
empty.
The batteries are flat.
The batteries have been
inserted incorrectly.
The cuff line is not correctly
inserted into the device.
The cuff is ruptured.
The cuff was not attached
correctly.
You moved or talked during
the measurement.
The measurement was ob-
structed by clothing.
53
Solution
Repeat the measurement as
perthe instructions in these
Instructions for use.
Replace the batteries.
Replace the batteries.
Re-insert the batteries, observ-
ing the correct polarity (-/+).
Ensure that the cuff line is cor-
rectly inserted into the device.
Replace the cuff. Contact
Customer Services.
Re-attach the cuff.
Do not move or talk during the
measurement.
Ensure that clothing cannot
obstruct the arm cuff during the
measurement.

Publicidad

Tabla de contenido
loading

Tabla de contenido