11.3 Scanning the image data via a
computer
Starting the imaging plate scanner and soft-
ware
The scanning process using the DBSWIN
imaging software is described
For further information on using the imag-
ing software, refer to the relevant manual
Press the on/off switch
i
unit
Switch on the computer and monitor
i
Start DBSWIN
i
Select the patient
i
Select the exposure values in the X-ray mod-
i
ule
Set the required resolution
i
Click the Scan button
i
If ScanManager is enabled, select the X-ray
i
job on the touch screen of the unit
Result:
An animation on the touch screen prompts for
insertion of the imaging plate
Insert the imaging plate only once the bar
above the animation is green
Figure 7: Example of the animation prompting
for insertion of the imaging plate
2151100018L40 1903V012
OBJ_DOKU-0000003578-mehrere-011.indd 37
Scanning an intraoral phosphor storage plate
i
to switch on the
i
i
i
CAUTION
Loss of image data caused by stray
ambient light entering the unit
Remove the plate guide only after the
i
phosphor storage plate has dropped
into the collecting tray
Slide the phosphor storage plate into an un-
occupied plate guide until the PSP is auto-
matically drawn in
The inactive side must be visible
Scanning progress is displayed on the touch
screen Image data is saved automatically
The touch screen only shows a pre-
view as an initial impression of the X-
ray image Limitations to image pre-
views occur due to image size and/or
exposure conditions For purposes of
diagnosis, the X-ray image must be
viewed on a diagnostic monitor
After it has been scanned, the phosphor
storage plate is erased and drops into the
collecting tray
Save the X-ray image
Remove the PSP and prepare it for recording
another X-ray image
Remove the plate guide after the entry slot
status LED on the touch screen switches to
green
Usage
EN
US
37
09.04.2019 10:51:18