en
Damage to the product due to inap-
propriate cleaning/disinfecting
CAUTION
agents!
Only use cleaning/disinfecting
►
agents approved for surface clean-
ing. Follow the manufacturer's
instructions for the respective
cleaning/disinfecting agent.
Never immerse the product in water
►
or detergent.
Do not introduce any fluid into the
►
charging bays.
Clean the contacts in the charging
►
bay, applying extreme care.
►
Wipe the shell of the product with a lint-free cloth moistened with a
commercial disinfectant for wipe disinfection.
►
If necessary, after the appropriate exposure time, wipe off any residues
of the cleaning/disinfecting agent with a lint-free cloth moistened
with clean water.
►
Dry the product with a clean, lint-free cloth.
►
Clean the contacts in the charging bay with isopropanol or ethyl alco-
hol and a cotton swab. Do not use corrosive chemicals.
►
If necessary, repeat the cleaning/disinfecting process.
6.2
Inspection and checking
►
Allow the product to cool down to room temperature.
►
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
►
Check the product for any damage, abnormal running noise, overheat-
ing or excessive vibration.
►
Inspect the shearing head for broken, damaged or blunt cutting edges.
►
Set aside the product if it is damaged.
7.
Maintenance
To ensure reliable operation, Aesculap recommends servicing the product
at least once a year.