Pressure accuracy
Maximum static pressure variation at 10 cm H
ISO 80601-2-70:2015
Without humidification
With humidification
Maximum dynamic pressure variation according to ISO 80601-2-70:2015
Device without humidification and Standard air tubing / Device with humidification and Standard air tubing
Pressure [cm H
O (hPa)]
2
4
8
12
16
20
Device without humidification and SlimLine air tubing / Device with humidification and SlimLine air tubing
Pressure [cm H
O (hPa)]
2
4
8
12
16
20
Flow (maximum) at set pressures
The following are measured accordingly to ISO 80601-2-70:2015 at the end of the specified air tubing:
Pressure
cm H
O (hPa)
2
4
8
12
16
20
Symbols
The following symbols may appear on the product or packaging.
Read instructions before use.
use.
Manufacturer.
Catalog number.
Protected against finger sized objects and against dripping water when tilted up to 15 degrees
from specified orientation.
equipment.
Humidity limitation.
China pollution control logo 1.
(In the US, Federal law restricts these devices to sale by or on the order of a physician).
20
Standard air tubing
± 0.5 cm H
O (± 0.5 hPa)
2
± 0.5 cm H
O (± 0.5 hPa)
2
10 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
10 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
AirSense 10 and
AirSense 10,
Standard
humidification and
Standard
L/min
L/min
180
143
168
135
157
136
144
134
131
123
Indicates a warning or caution.
European Authorized Representative.
Serial number.
O (10 hPa) according to
2
15 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
15 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
Device number.
Direct current.
Type BF applied part.
Temperature limitation.
China pollution control logo 2.
SlimLine air tubing
± 0.5 cm H
O (± 0.5 hPa)
2
± 0.5 cm H
O (± 0.5 hPa)
2
20 BPM
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
20 BPM
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
AirSense 10 and
SlimLine
L/min
162
151
140
128
117
Follow instructions before
Batch code.
On / Off.
Non-ionising radiation.
AirSense 10,
humidification and
ClimateLineAir
L/min
151
142
135
121
109
Device weight.
Class II
Prescription only