Flow (Maximum) At Set Pressures; Displayed Values - ResMed S9 Serie Manual Clínico

Equipo de tratamiento con presión positiva en las vías respiratorias
Ocultar thumbs Ver también para S9 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Flow (maximum) at set pressures

The following are measured at the end of the specified air tubing:
Pressure
cm H
O
2
4
8
12
16
20
25

Displayed values

Value
Pressure sensor at air outlet
Mask pressure
Flow derived values
Leak
Tidal volume
Respiratory rate
Minute ventilation
Ti
I:E ratio
Value
Pressure measurement
Mask pressure
Flow measurements
Leak
2
Tidal volume
2
Respiratory rate
2
Minute ventilation
1
Results are expressed at ATPD (Ambient Temperature and Pressure, Dry)
2
Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation
<3 L/min.
30
S9 VPAP Tx and
standard air
tubing,
L/min
200
200
200
200
190
180
Range
3–30 cm H
0–200 L/min
0–4000 mL
0–50 BPM
0–30 L/min
0.1–4.0 sec
1:50–2:1
Accuracy
1
±0.5 cm H
1
±12 L/min or 20% of reading, whichever is greater, at 0 to 60 L/min
±20%
±1.0 BPM
±20%
2
S9 VPAP Tx, H5i
and standard air
tubing,
L/min
170
170
170
170
170
161
O
2
1
O (+ 4% of measured value)
2
S9 VPAP Tx and
S9 VPAP Tx, H5i
SlimLine,
and ClimateLine,
L/min
195
190
184
175
168
144
Display resolution
0.1 cm H
O
2
1 L/min
1 mL
1 BPM
0.1 L/min
0.1 sec
0.1
L/min
170
170
170
170
161
125

Hide quick links:

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

S9 vpap t

Tabla de contenido