Errors and warnings

3.4.3

Replacing the alkaline batteries

WARNING
!
Do not remove or replace the alkaline batteries or the alkaline
pack in areas subject to possible explosion. Only use alkaline
batteries that are indicated on the alkaline pack, or else the
approval will become invalid and there is a risk of explosion.
1. Remove the 2 screws (1) with an appropriate hex key.
2. Pull the alkaline pack out of the Dräger X-act 5000.
3. Remove the lid (3) from the alkaline pack.
4. Replace the alkaline batteries with new ones. Be sure to observe the
polarity of the batteries (marked inside the compartment (2) of the
alkaline pack).
5. Put the lid back on the alkaline pack.
6. Insert the alkaline pack into the Dräger X-act 5000.
7. Fasten the 2 screws with an appropriate hex key.
20
4
Errors and warnings
4.1

Errors

Errors will be displayed together with the corresponding icon and a
code number or a written hint. The red LEDs will flash until
acknowledged with the "OK" key.
Text/Code
Cause
LEAK TEST
SO
ERROR
correctly in the
instrument; or air path in
the instrument is leaky
SCAN
Bar code on the tube box
1)
is unknown (e. g. when
ERROR
scanning a new tube).
Bar code on the tube box
is destroyed
Reader is damaged
TUBE
Inserted tube or sampling
BLOCKED
system is clogged or air
3
path in the instrument is
clogged.
BATTERY
Rechargeable battery is
ERROR
deeply discharged or the
2
type of charger is not
correct.
Rechargeable battery is
1
defective
00321900.eps
Remedy
filter does not fit
Repeat leak test. If leak test
3
failed, check the SO
for correct seat or contact
DrägerService
Set strokes manually or
update tube database using
password- protected menu.
Set strokes manually.
Contact DrägerService.
Use new tube or sampling
system or check that SO
filter fits tight or contact
DrägerService.
Wait for 1 minute and then
clear the error by pressing
the "OK" key or use correct
charger or contact
DrägerService.
Replace rechargeable
battery.
filter
3
®
.
3
Dräger X-act 5000