Technical Data - Seca 674 Manual De Instrucciones

Ocultar thumbs Ver también para 674:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

E
... the display
and a number appear?
– Switch off the scale with the Start key and start the scale again. The scale will then
work normally again.
If this is not the case, disconnect the power supply by briefly removing the mains unit.
If this measure is equally unsuccessful, inform the Service department.
9. Servicing
Your seca scale leaves the factory with an accuracy greater than ±0.15 %. To ensure that
this accuracy is maintained, the product must be carefully set up and regularly maintained.
Depending on the frequency of use, we recommend servicing every 3 to 5 years. Please
ensure that a qualified service agent is used. If in doubt, contact seca Customer Service for
further information.

10. Technical data

Dimensions scale
width:
height:
depth:
Dimensions weighing platform
width:
height:
depth:
Height of figures
Weight
Temperature range
EN 60 601-1:
Insulated equipment, Protection Class II:
medical electrical equipment type B:
Colour
Maximum load
Graduations
Accuracy
Power supply
24
890 mm / 35 ins
+10
+0.4
65
mm / 2.6
965 mm / 38 ins
800 mm / 31.5 ins
+10
+0.4
46
mm / 1.8
965 mm / 38 ins
20 mm
approx. 25.5 kg
+10 °C to +40 °C
silver grey/black
360 kg / 800 lbs / 57 sts
100 g / 0.2 lbs (USA: 0.1 lbs)
greater than ±0.15 % / ±100 g
mains unit
ins
ins

Publicidad

Tabla de contenido
loading

Tabla de contenido