4 Operating

4.1

Use and Function

14
1. Familiarise yourself with the safety notes in Sections 1.1 and 1.2 on pages
10 – 11 and then proceed as described in Section 3.1 points 1 – 5.
2. Place electrode pad
with firm pressure. Switch the device on by pressing ON/ + button
Function display
the device starts to work at the lowest level of intensity. The device has 10
levels of intensity for strength of treatment ranging from 0 - 60 mA. Select
these intensity levels by simply pressing ON/ + button
number of times.
3. Your device has an automatic cut-out function which switches it off auto-
matically after 20 minutes of treatment.
4. A treatment program with three different pulse patterns runs during the
preset 20-minute treatment period.
Pulse pattern 1:
Pulse frequency 30 - 60 Hz
Pulse width 250 milliseconds
Pulse frequency varies automatically in a 6 second on, 1 second off rhythm
Pulse pattern 2:
Pulse frequency 30 Hz
Pulse width 150 - 250 milliseconds
Pulse frequency varies automatically in a 6 second on, 1 second off rhythm
Pulse pattern 3:
Pulse frequency 30 - 60 Hz
Pulse width 150 - 250 milliseconds
Pulse frequency varies automatically in a 6 second on, 1 second off rhythm
The three pulse patterns run consecutively for 1 minute 20 seconds each.
This pattern is repeated 5 times during the 20-minute treatment period. The
device's operation is displayed visually in that function display
with decreased brilliancy in line with the rhythm of the treatment program.
5. Treatment may be discontinued at any time by pressing - /OFF button
repeatedly (depending on the level of intensity) until red function display
lights up more brightly again. Now press button
device is switched off.
6. The adhesion of electrode pad
times. Therefore, take great care when removing the electrode pad and
device from the skin. Never pull on the device. After use, replace the
electrode pad on the protective plastic film for storage.
on the area of skin selected and smooth it down
lights up brightly. By pressing ON/ + button
is very strong when used the first few
.
again,
the appropriate
flashes
one more time and the