Cleaning the analyzer
To clean the case and screen (perform as needed)
Clean the case and screen using a lint-free cloth dampened with 70% isopropyl alcohol, mild liquid soap and water,
or a 10% bleach solution. DO NOT wipe the vents underneath the power button or the USB/power/Ethernet ports.
To clean the internal components (perform weekly or as needed)
1. Power off the analyzer
and disconnect the
power and/or Ethernet
cord.
2. Use a flat-tip screw
driver or coin to turn
the 2 silver fasteners
counterclockwise a
quarter turn.
3. Remove the base from
the analyzer.
4. Clean the entire base
using a lint-free wipe or
cotton swab dampened
with 70% isopropyl
alcohol, mild liquid soap
and water, or a 10%
bleach solution.