Calibrating the Sensors
EN
This section describes how to use the Cyclocomputer SGX-CA500 to calibrate
the zero point of the pedaling monitor sensor that is installed on the bicycle.
Zero point calibration is a function to store the zero point (no-load), where no
forces act on the crank, in the sensor memory.
•
The right-side pedaling monitor sensor is used as an example in this description. The
procedure to calibrate the left side is the same as for the right side.
•
Please do not push the push switch in the right transmitter while calibrating the sensor
or showing [Force Preview] with the Cyclocomputer SGX-CA500.
Getting Ready
1.
Stop the bicycle on a flat safe
place.
Calibrating the Zero Point
1.
Position the crank arm so it is
perpendicular to the ground,
pointing downward.
14
2.
Tap the [Sensors] icon in the
home screen of the SGX-
CA500.
The sensor list screen opens.
3.
Tap [Pedaling Monitor R] and
then [Calibration (Zero)].
4.
Tap [Start Calibration].
The calibration starts.
If the calibration is successful,
"Success" appears in the [Result] field.
If "Failure" is displayed, the sensor may
be calibrated in an unstable condition
causing the crank to be moving during
the calibration. Calibrate again with the
crank stopped.