Electromagnetic Compatibility Information

Table 1
Guidance and manufacturer's declaration – electromagnetic emissions
The BC 30 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 30 should assure that it
is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Table 2
Guidance and manufacturer's declaration – electromagnetic immunity
The BC 30 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 30 should assure that it
is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
Compliance
Electromagnetic environment – guidance
Group 1
The BC 30 uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
Class B
The BC 30 is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that sup-
Not applicable
plies buildings used for domestic purposes.
Not applicable
IEC 60601 test level Compliance level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30 %.
Power frequency magnetic fields should be at levels characteris-
tic of a typical location in a typical commercial or hospital envi-
ronment.
86