Recommended separation distances between portable and mobile RF communications equipment and the BC 30
The BC 30 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user
of the BC 30 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communica-
tions equipment (transmitters) and the BC 30 as recommended below, according to the maximum output power of the communications equip-
ment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structu-
res, objects and people.
Beurer GmbH
Söflinger Str. 218
www.beurer-healthguide.com
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
Not applicable
89077 Ulm, Germany
www.beurer-gesundheitsratgeber.com
88
(m)
80 MHz to 800 MHz
d = 1.2
0.12
0.38
1.2
3.8
12
www.beurer.com
800 MHz to 2.5 GHz
d = 2.3
0.23
0.73
2.3
7.3
23