Preparations at the place of use

If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example.
Remove any visible surgical residues to the extent possible with a damp, lint-free cloth.
Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.

Preparation before cleaning

Dismantle the product prior to cleaning, see Disassembling.
Cleaning/disinfection
Product-specific safety notes on the reprocessing procedure
Damage to the product due to inappropriate cleaning/disinfecting agents and/or
excessive temperatures!
Use cleaning and disinfecting agents according to the manufacturer's instruc-
tions which
CAUTION
– are approved for the material in question (e.g., aluminum, plastics, high-
grade steel),
Observe specifications regarding concentration, temperature and exposure
time.
Do not exceed the maximum permitted cleaning temperature of 93 °C.
Immersion treatment in a 3 % H
O
solution for approx. 5 minutes is a particularly effective and gentle method to
2
2
dissolve encrustations from HF instruments. Subsequently, the debris can be removed by hand, with a medium-hard
brush and or in an ultrasonic bath. This is followed by the conventional reprocessing steps.
Use suitable cleaning/disinfecting agents if the product is put away in a wet condition. To prevent foam forma-
tion and reduced effectiveness of the process chemicals: Prior to mechanical cleaning and disinfection, rinse the
product thoroughly with running water.
Carry out ultrasound cleaning:
– as an effective mechanical supplement to manual cleaning/disinfection.
– as a pre-cleaning procedure for products with encrusted residues, in preparation for mechanical cleaning/
disinfection.
– as an integrated mechanical support measure for mechanical cleaning/disinfection.
– for additional cleaning of products with residues left after mechanical cleaning/disinfection.
Validated cleaning and disinfection procedure
Validated procedure
Specific requirements
Manual cleaning with immersion
Cleaning brush, e.g., GK469200
disinfection
20 ml disposable syringe
Handle
When cleaning products with
movable hinges, ensure that
these are in an open position
and, if applicable, move the
joint while cleaning.
Drying phase: Use a lint-free
cloth or medical compressed air
Manual cleaning with ultrasound
Cleaning brush, e.g., GK469200
and immersion disinfection
20 ml disposable syringe
Shafts
When cleaning instruments
with movable hinges, ensure
that these are in an open posi-
tion and, if applicable, move
the joint while cleaning.
Drying phase: Use a lint-free
cloth or medical compressed air
Mechanical alkaline cleaning and
Place the instrument in a tray
thermal disinfection
that is suitable for cleaning
(avoiding rinsing blind spots).
Handle
Connect components with
lumens and channels directly to
the rinsing port of the injector
carriage.
Keep working ends open for
cleaning.
Place instruments in the tray
with their hinges open.
Manual pre-cleaning with brush
Cleaning brush, e.g., GK469200
and subsequent mechanical alka-
20 ml disposable syringe
line cleaning and thermal disinfec-
tion
Place the instrument in a tray
that is suitable for cleaning
Shafts
(avoiding rinsing blind spots).
Connect components with
lumens and channels directly to
the rinsing port of the injector
carriage.
Keep working ends open for
cleaning.
Place instruments in the tray
with their hinges open.
Reference
Chapter Manual cleaning/disinfec-
tion and sub-chapter:
Chapter Manual cleaning with
immersion disinfection
Chapter Manual cleaning/disinfec-
tion and sub-chapter:
Chapter Manual cleaning with
ultrasound and immersion dis-
infection
Chapter Mechanical cleaning/dis-
infecting and sub-chapter:
Chapter Mechanical alkaline
cleaning and thermal disinfect-
ing
Chapter Mechanical cleaning/dis-
infection with manual pre-clean-
ing and sub-chapter:
Chapter Manual pre-cleaning
with a brush
Chapter Mechanical alkaline
cleaning and thermal disinfect-
ing
Manual cleaning/disinfection
Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution.
After manual cleaning/disinfection, check visible surfaces visually for residues.
Repeat the cleaning /disinfection process if necessary.
Manual cleaning with immersion disinfection
Phase
Step
T
[°C/°F]
I
Disinfecting
RT (cold)
cleaning
II
Intermediate
RT (cold)
rinse
III
Disinfection
RT (cold)
IV
Final rinse
RT (cold)
V
Drying
RT
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)
RT:
Room temperature
*Recommended: BBraun Stabimed
Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.
Phase I
Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are
moistened.
Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.
If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.
Mobilize non-rigid components, such as set screws, links, etc. during cleaning.
Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible surfaces) under running water.
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Drain any remaining water fully.
Phase III
Fully immerse the product in the disinfectant solution.
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe.
Ensure that all accessible surfaces are moistened.
Phase IV
Rinse/flush the product thoroughly (all accessible surfaces).
Mobilize non-rigid components, such as set screws, joints, etc. during final rinse.
Rinse lumens with an appropriate disposable syringe at least five times.
Drain any remaining water fully.
Phase V
Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning
and disinfection procedure.
Manual cleaning with ultrasound and immersion disinfection
Phase
Step
T
[°C/°F]
I
Ultrasonic
RT (cold)
cleaning
II
Intermediate
RT (cold)
rinse
III
Disinfection
RT (cold)
IV
Final rinse
RT (cold)
V
Drying
RT
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)
RT:
Room temperature
*Recommended: BBraun Stabimed
Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.
Phase I
Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible
surfaces are immersed and acoustic shadows are avoided.
Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.
If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.
Mobilize non-rigid components, such as set screws, links, etc. during cleaning.
Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible surfaces) under running water.
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Drain any remaining water fully.
t
Conc.
Water
Chemical
[%]
quality
[min]
>15
2
D–W
Aldehyde-free, phenol-free,
and QUAT-free concentrate,
pH ~ 9*
1
-
D–W
-
15
2
D–W
Aldehyde-free, phenol-free,
and QUAT-free concentrate,
pH ~ 9*
1
-
FD-W
-
-
-
-
-
t
Conc.
Water
Chemical
[%]
quality
[min]
>15
2
D–W
Aldehyde-free, phenol-free,
and QUAT-free concentrate,
pH ~ 9*
1
-
D–W
-
15
2
D–W
Aldehyde-free, phenol-free,
and QUAT-free concentrate,
pH ~ 9*
1
-
FD-W
-
-
-
-
-