Hose with screw connection
 Screw hose with screw connection M 12 x 1 (A)
firmly by hand onto O
connection M 12 x 1.
2
An unused O
connection will close automatical-
2
ly.
Hose with coupling plug
Instructions for use Alduk IV
 Insert coupling plug firmly into quick coupling
connection.
If no O
flow from the nozzle is needed:
2
A
B
 Turn the selector wheel (B) to position 0 until it
locks into place.
 Open cylinder valve slowly for at least one full
revolution.
Using the additional O
Examples for using the additional O
Filling a manual breathing bag with O
Using for inhalation or insufflation
WARNING
Do not use the O
2
nector to drive medical devices.
WARNING
Always ensure that the selector wheel is in a
locked position. In an intermediate position, the
device will supply no flow.
1
2
flow
2
flow:
2
The accuracy of the flow depends on the total
2
amount of O
the same time a high amount of O
another O
maintained.
connection with nozzle con-

After use

1
2
B
A
Fit the hose (ø 5 mm) of the connected
device (A) onto nozzle.
Set the O
flow at the selector wheel (B).
2
The following O
flow rates can be selected: 1, 2,
2
3, 4, 6, 8, 12, 15 L/min.
withdrawn at the pressure reducer. If at
2
is withdrawn from
2
connection, the set O
flow cannot be
2
2
Close the cylinder valve completely.
Relieve the pressure reducer via the connected
device until the pressure gauge displays 0 bar.
English
15