DESCRIPTION OF OPERATION
The biological incubator is a device for use in medical and/or dental.
The producer is not responsible for damage to the unit and personal
injuries caused by improper or lack of experience of operator.
Temperature of usage +17 / +27 ° C
On incubator there is an indicator light to indicate the possible bad‐functioning of
the machinery or the momentary lack of electricity (and then the failure of tests in
progress). Also inside the device there is a thermostat security in case of reaching
temperatures than those using interrupts the functioning of resistance.
1.
Connect the power cord (fg.1 – POS.D) to the electrical line.
2.
The warning led (fg.1 – POS.A) lights.
3.
Push the START button (fg.1 – POS.E). The led switch off and the heating cycle
of unit start. When occurs a drop of temperature, caused by natural cooling of
the unit, the electronic card order to switch again the resistance. In case of
failure to reach temperature or sudden interruption of power supply, the led
stays on e the incubator stops working.
Reliability of the validity of the test
If at the end of incubation time (about 24 hours) the LED is off the test is valid.
If at the end of incubation time the LED is on, the test is NOT valid
CAUTION: HOT SURFACE – Do not touch the heating block during
operation of the unit.
CAUTION: Keep away from children
CAUTION: Do not touch metal surfaces when the machine is running