Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM –
for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the BC85 blood pressure monitor
The BC85 blood pressure monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are control-
led. The customer or the user of the BC85 blood pressure monitor can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the BC85 blood pressure monitor as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be esti-
mated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-
tures, objects and people.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d =
0.117
0.370
1.170
3.700
11.7
119
(m)
80 MHz to 800 MHz
d =
0.117
0.370
1.170
3.700
11.7
800 MHz to 2.5 GHz
d =
0.234
0.740
2.340
7.400
23.4