Maintenance And Cleaning; Accessories - IKA RV 10 control Instrucciones De Manejo

Ocultar thumbs Ver también para RV 10 control:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

PC 1.1 cable
This cable is required to connect the 9-pin connector to a PC.
1
6
2
7
3
8
4
9
5
1
RxD 2
TxD 3
4
GND 5
6
RTS 7
CTS 8
9
The device is maintenance-free. It is only subject to the natural
wear and tear of components and their statistical failure rate.
The seal on the glass condenser should be checked at regular in-
tervals and replaced if necessary.
Cleaning
• For cleaning disconnect the mains plug!
• Use only cleaning agents which have been approved by IKA to
clean IKA devices.
These are water (with tenside) and isopropanol.
• Wear protective gloves during cleaning the devices.
• Electrical devices may not be placed in the cleansing agent for
the purpose of cleaning.
• Do not allow moisture to get into the device when cleaning.
• Before using another than the recommended method for clean-
ing or decontamination, the user must ascertain with IKA that
this method does not destroy the device.
Spare parts order
When ordering spare parts, please give:
- device type.
- serial number, see type plate.
- position number and description of spare part, see
www.ika.com,
- software version
For further accessories see www.ika.com.
26
USB 2.0 cable
This cable is required to connect USB interface to a PC.
5
1
9
4
2
8
3
4
3
7
5
2
6
1
USB
1
2 RxD
3 TxD
PC
4
5 GND
6
7 RTS
8 CTS
9

Maintenance and cleaning

Repair
Please only send devices in for repair that have been cleaned
and are free of materials which might present health hazards.
For repair, please request the "Safety Declaration (Decontami-
nation Certificate)" from IKA or use the download printout of it
from IKA website at www.ika.com.
If your appliance requires repair, return it in its original packaging.
Storage packaging is not sufficient when sending the device - also
use appropriate transport packaging.
Note: For glassware unit, please contact your local dealer and do
not send the glassware back to our factory.
Safety lift
The safety lift must be checked daily prior to operation!
After a long period of inactivity (approx. four weeks), use the mo-
tor to move the lift to the top or bottom position several times
prior to starting distillation. (See the section entitled Safety In-
structions - Safety Lift)! If the safety lift is not working, please
contact the IKA Service department.

Accessories

1
2
3
4
USB

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Rv 10 auto

Tabla de contenido