Spectrum For Photo Detector; Measuring Procedure - KPS LX10 Manual De Funcionamiento

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

KPS-LX10 • Digital light meter
3-3 Correction Factor
• Mercury Lamp . ................................... x1.1
• Fluorescent Lamp .............................. x1.0
• Incandescent Light............................. x1.0
• Daylight . ............................................. x1.0
4.

SPECTRUM FOR PHOTO DETECTOR

5.

MEASURING PROCEDURE

• Power-up: Press the POWER key to turn the meter on.
• Selecting the LUX scale: Set the Function switch to desired LUX range.
• Face the photo detector to light source in a horizontal position.
• Read the illuminance value from the LCD display.
• Data-Hold mode: Press the HOLD key to select Hold mode. When
• HOLD mode is selected, the illuminance meter stops all further measure-
ments. Press the HOLD key again to cancel HOLD mode.
• Then it resumes normal operation.
• When the measurement is completed, turn the power off.
Note:
If display indicates one or more leading zeros, user has toshift the Function
switch to the next lower range scale to improve resolution and accuracy.
6.
MAINTENANCE
6-1 General maintenance
• To keep the instrument clean, wipe the case with a dry cloth and detergent,
do not use abrasives or solvents.
• The white plastic disc on the top of the detector should be cleaned with a
damp cloth when necessary.
100
80
60
40
20
400
500
600
700
Longitud de onda (nm)
9
800
900
1000

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

602450002

Tabla de contenido