SETUP
After connecting to the 2
weighing platform, go to menu setup to set the 2
nd
Plat.Cap.Unit and 2nd Plat.Grad and then perform a calibration.
2nd Plat.Capacity
Set the Capacity of Scale 2 using the numerical keypad
3 kg
...
30 kg
...
10000 kg
2nd Plat.Cap.Unit
Use this menu to set the primary weighing unit for Scale2:
Kilogram
Pound
2nd Plat. Grad.
Use this menu to set the readability of scale 2
0.0005
0.001
0.002
0.005
0.01
0.02
Note: When the scale2 is used in Legal for trade, Resolution <= 3000e.
Press and hold the
key on the overlay to switch between the main platform and the secondary
platform.
Note: Make sure to perform a calibration of the scale before using the 2
www.
2
Scale Interface
nd
Plat.Capacity, 2
nd
nd
weighing platform.