Metrohm 784 KFP Titrino Instrucciones Para El Uso página 73

Ocultar thumbs Ver también para 784 KFP Titrino:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

3.2 Remote control commands
68
The limit control for results can also be activated. If a result is out of limit, a
message appears in the result report, E196 is sent, and output line L13 can
be set.
3.2.2.6.
Mode.Def.SiloCalc.Assign.C24
Mode.Def.SiloCalc.Assign.C25
Mode.Def.SiloCalc.MatchId
.Assign.C2X: Assignment to store results in the silo as C2X.
.MatchId:
Indication which sample identification(s) have to match so that
the results can be combined.
3.2.2.7.
Mode.Def.ComVar.C30
Mode.Def.ComVar.C31
etc., up to .C39
Assignment of common variables.
The values of the common variables are to be found in &Config.ComVar. They
can be viewed and entered there, see 3.2.2.50.
3.2.2.8.
Mode.Def.Report.Assign1
Mode.Def.Report.Assign2
param, full, short, mplist, curve, scalc full, scalc srt, calc, ff
Definition of the report sequence, which is outputted automatically at the end
of the determination. Entries of more than one block have to be separated
with ";".
.Assign1: Output to COM1 of the Titrino. Identical for COM2.
3.2.2.9.
Mode.Def.Mean.1.Assign
Mode.Def.Mean.2.Assign
etc., up to .9
Assignment of the statistics calculations. Valid assignments are a require-
ment for statistics calculations. In addition, the statistics calculation must be
switched on, see 3.2.2.24. Rules for statistics calculations see page 29.
3.2.2.10. Mode.CFmla
Mode.CFmla.1.Value
Mode.CFmla.2.Value
etc., up to .19
Calculation constants specific to a method. Stored in the method memory of
the Titrino. Operands specific to the sample (3.2.2.52 and 3.2.2.63) and val-
ues of common variables (3.2.2.50) on the other hand are not stored with the
methods.
RSX, EPX, CXX
RSX, EPX, CXX
id1, id1&2, all, OFF
RSX, MNX, EPX, CXX
RS1, RSX, EPX, CXX
0...±999999
784 KFP Titrino

Publicidad

Tabla de contenido
loading