784 KFP Titrino
3.2.2.47. Config.Aux.DevName
Name of the instrument for connections with several units. It is advisable to
use only the letters A...Z (ASCII No. 65...90), a...z (ASCII No. 97...122) and
the numbers 0...9 (ASCII No. 48...57) when the function Setup.AutoInfo
(3.2.2.92) is used at the same time.
If a name has been entered, it will be printed out in the result report (full,
short).
3.2.2.48. Config.Aux.Prog
Output of the program version.
The Titrino sends "784.0010" on requests with $Q.
3.2.2.49. Config.RSSet1
Config.RSSet1.Baud
Config.RSSet1.DataBit
Config.RSSet1.StopBit
Config.RSSet1.Parity
Config.RSSet1.Handsh
$G sets all RS settings. The changes are performed only if the instrument is
inactive. After the setting of the interface parameters, wait at least 2 s to allow
the components to equilibrate.
Settings of the values for the data transmission via the RS interface: baud
rate, data bit, stop bit, parity and type of handshake, see also page 90ff. Baud
rates >9600 need a PC which is equipped accordingly (e.g. with 16550
component).
The setting of the values must be initiated with $G immediately after entry of
the values.
3.2.2.50. Config.ComVar.C30
with up to .C39, etc.
Values of the common variables from C30 up to C39. Insert the common va-
riables directly or describe the determination results directly from the method,
see 3.2.2.7.
3.2.2.51. SmplData.Status
On/off switching of silo memory. When the silo memory is switched on, the
sample data are fetched from the lowest valid silo line.
3.2 Remote control commands
up to 8 ASCII characters
300, 600, 1200, 2400, 4800, 9600,
19200, 38400, 57600, 115200
even, odd, none
HWs, SWchar, SWline, none
0... ±999999
read only
$G
7, 8
1, 2
ON, OFF
75